5

PROFIBUS SETTINGS

46

UFC 400

www.krohne.com

05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en

5.3 Menu C, Setup

if condensed status and diagnosis

condensed status and diagnosis

condensed status and diagnosis

condensed status and diagnosis handling selected:

No.

Function

Settings / descriptions

C1 Process Input

C1.10

Diagnostics

-

It is strongly recommended to use the default classification of the variable "Event groups" mentioned below. If the

default classification should be changed please note!

1

C1.10.5

Proc: Signal Unreliable

Select:Out of Specification (default)FailureInformationMaintenance

RequiredFunction Check

1

C1.10.3

Proc: Empty Pipe

Select:Out of Specification (default)FailureInformationMaintenance

RequiredFunction Check

1

C1.10.4

Proc: Signal Lost

Select:Out of Specification (default)FailureInformationMaintenance

RequiredFunction Check

1

C1.10.6

Config: Totaliser

Select:Out of Specification (default)FailureInformationMaintenance

RequiredFunction Check

1

C1.10.8

Electr: Power Failure

Select:Information (default)Maintenance RequiredFunction CheckOut of

SpecificationFailure

1

C1.10.7

Electr: IO Connection

Select:Out of Specification (default)FailureInformationMaintenance

RequiredFunction Check

1

1 Below mentioned definitions for both diagnosis and status will be internally used for PROFIBUS diagnosis and status handling of the

classifications stated below which will be possible for the variable "Event groups":

"F"

Failure

diagnosis: DIA_MAINTENANCE_ALARM (Maintenance alarm (40))

status: BAD - maintenance alarm

"S"

Out of specification

diagnosis: DIA_MAINTENANCE_DEMAND (Maintenance demanded (41))

status: UNCERTAIN - process related, no maintenance

"M"

Maintenance required

diagnosis: DIA_MAINTENANCE (Maintenance required (37))

status: GOOD - maintenance demanded

"C"

Checks in progress

diagnosis: DIA_FUNCTION_CHECK (Function check (42))

status: GOOD - function check

"I"

Information

diagnosis: NO DIAGNOSIS EVENTS (no diagnosis events)

1

status: GOOD - ok

1 Bit "EXTENSION_AVAILABLE" of parameter DIAGNOSIS will be set if at least one Bit of parameter "DIAGNOSIS_EXTENSION" has been

set.