PROFIBUS SETTINGS
5
51
UFC 400
www.krohne.com
05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en
5.5 Status messages and diagnostic information
Error
type
Event group
Single event
Description
Actions to eliminate the event
F
F
F
F
F Electronics
PROFIBUS Failure
Error in the electronics due to an
internal bus communication
timeout or a hardware error.
Perform a power reset. If the
status returns, contact the
factory.
F
F
F
F
F Configuration
PROFIBUS Config.
Invalid PROFIBUS configuration.
Check PROFIBUS configuration
or reset to factory configuration
(D2.2.8).
Tot 1 FB3 Unit Error
The totaliser is not counting,
because the unit is not known.
Check unit in totaliser 1 FB3 or
reset to factory settings (D2.2.8).
Tot 2 FB4 Unit Error
Check unit in totaliser 2 FB4 or
reset to factory settings (D2.2.8).
Tot 3 FB5 Unit Error
Check unit in totaliser 3 FB5 or
reset to factory settings (D2.2.8)
C
C Configuration
PROFIBUS Sim/Man
Active
A simulation function in the
PROFIBUS component is active
and in use.
Switch off PROFIBUS simulation.
Tot 1 FB3 Sim. Active
A simulation function in the
PROFIBUS component is active
and in use.
(for internal testing purpose
only)
Switch off PROFIBUS simulation.
Tot 2 FB4 Sim. Active
A simulation function in the
PROFIBUS component is active
and in use.
(for internal testing purpose
only)
Switch off PROFIBUS simulation.
Tot 3 FB5 Sim. Active
A simulation function in the
PROFIBUS component is active
and in use.
(for internal testing purpose
only)
Switch off PROFIBUS simulation.
Tot 1 FB3 Man. Active
The manual mode in the
PROFIBUS component is active.
Switch off PROFIBUS manual
mode.
Tot 2 FB4 Man. Active
The manual mode in the
PROFIBUS component is active.
Switch off PROFIBUS manual
mode.
Tot 3 FB5 Man. Active
The manual mode in the
PROFIBUS component is active.
Switch off PROFIBUS manual
mode.
S
S Electronics
Electr.Temp. Out of
Spec
The electronics temperature
measured inside the electronics
housing exceeds the limits.
Move electronics away from
process influences. Protect the
electronics against direct
sunlight.