beautypg.com

Profibus settings, 5 status messages and diagnostic information – KROHNE UFC 400 PROFIBUS PA-PROFIBUS DP User Manual

Page 51

background image

PROFIBUS SETTINGS

5

51

UFC 400

www.krohne.com

05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en

5.5 Status messages and diagnostic information

Error
type

Event group

Single event

Description

Actions to eliminate the event

F

F

F

F

F Electronics

PROFIBUS Failure

Error in the electronics due to an

internal bus communication

timeout or a hardware error.

Perform a power reset. If the

status returns, contact the

factory.

F

F

F

F

F Configuration

PROFIBUS Config.

Invalid PROFIBUS configuration.

Check PROFIBUS configuration

or reset to factory configuration

(D2.2.8).

Tot 1 FB3 Unit Error

The totaliser is not counting,

because the unit is not known.

Check unit in totaliser 1 FB3 or

reset to factory settings (D2.2.8).

Tot 2 FB4 Unit Error

Check unit in totaliser 2 FB4 or

reset to factory settings (D2.2.8).

Tot 3 FB5 Unit Error

Check unit in totaliser 3 FB5 or

reset to factory settings (D2.2.8)

C

C Configuration

PROFIBUS Sim/Man

Active

A simulation function in the

PROFIBUS component is active

and in use.

Switch off PROFIBUS simulation.

Tot 1 FB3 Sim. Active

A simulation function in the

PROFIBUS component is active

and in use.

(for internal testing purpose

only)

Switch off PROFIBUS simulation.

Tot 2 FB4 Sim. Active

A simulation function in the

PROFIBUS component is active

and in use.

(for internal testing purpose

only)

Switch off PROFIBUS simulation.

Tot 3 FB5 Sim. Active

A simulation function in the

PROFIBUS component is active

and in use.

(for internal testing purpose

only)

Switch off PROFIBUS simulation.

Tot 1 FB3 Man. Active

The manual mode in the

PROFIBUS component is active.

Switch off PROFIBUS manual

mode.

Tot 2 FB4 Man. Active

The manual mode in the

PROFIBUS component is active.

Switch off PROFIBUS manual

mode.

Tot 3 FB5 Man. Active

The manual mode in the

PROFIBUS component is active.

Switch off PROFIBUS manual

mode.

S

S Electronics

Electr.Temp. Out of

Spec

The electronics temperature

measured inside the electronics

housing exceeds the limits.

Move electronics away from

process influences. Protect the

electronics against direct

sunlight.