Profibus settings, 4 menu d, service – KROHNE UFC 400 PROFIBUS PA-PROFIBUS DP User Manual
Page 50
5
PROFIBUS SETTINGS
50
UFC 400
www.krohne.com
05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en
5.4 Menu D, Service
This menu is protected. You will need to use the service password to gain access.
C5.8.4
Diag. extension h
h = history
Content of the PROFIBUS diagnosis extension history is displayed. Shows
PROFIBUS diagnosis extension information of all PROFIBUS diagnosis
extension bits which are set since last reset (all bits reset to zero – use
service menu D2.2.9 to reset this parameter)
C5.8.5
Diag. extension ih
h = history
Content of the internal PROFIBUS diagnosis extension history is displayed.
Shows internal PROFIBUS diagnosis extension information of all internal
PROFIBUS diagnosis extension bits which are set since last reset (all bits
reset to zero – use service menu D2.2.10 to reset this parameter)
C5.8.6
Identification No.
Sets different device modes for the cyclic configuration of the PROFIBUS
interface.
Select:
UFC400: all manufacturer specific parameters / functions supported
Profile: profile defined parameters / functions supported only
Autom.: (default setting) PROFIBUS configuration (UFC400 or Profile) will be
selected automatically by the PROFIBUS master itself without interaction by
the user. The active selection is configured during startup by using the
Set_Prm and/or Set_Slave_Add service. No additional acyclic parameter
settings required.
C5.8.8
Condensed Status
Defines the general method how the device handles status and diagnostics:
On: Condensed status and diagnosis information according PROFIBUS
Profile 3.02 will be used
Off: Classic status and diagnosis information according PROFIBUS Profile
3.02 will be used
No.
Function
Settings / descriptions
D4.0
Service Parameter
D4.7.0
Identification No.
Same as C5.8.6
D4.8.0
PB Cold Start
Initiates a PROFIBUS cold start.
During a PROFIBUS cold start nearly all parameter values of the whole
device will be set to their default values (Exceptions are the PROFIBUS
address and the identification No.). The PROFIBUS cold start will be carried
out without disconnecting an already established connection to a PROFIBUS
master system.
Select:
No (terminates this function) / Yes (PROFIBUS cold start will be carried out
at once)