5

PROFIBUS SETTINGS

50

UFC 400

www.krohne.com

05/2014 - 4003118502 - AD UFC 400 PROFIBUS R02 en

5.4 Menu D, Service

This menu is protected. You will need to use the service password to gain access.

C5.8.4

Diag. extension h

h = history

Content of the PROFIBUS diagnosis extension history is displayed. Shows

PROFIBUS diagnosis extension information of all PROFIBUS diagnosis

extension bits which are set since last reset (all bits reset to zero – use

service menu D2.2.9 to reset this parameter)

C5.8.5

Diag. extension ih

h = history

Content of the internal PROFIBUS diagnosis extension history is displayed.

Shows internal PROFIBUS diagnosis extension information of all internal

PROFIBUS diagnosis extension bits which are set since last reset (all bits

reset to zero – use service menu D2.2.10 to reset this parameter)

C5.8.6

Identification No.

Sets different device modes for the cyclic configuration of the PROFIBUS

interface.

Select:

UFC400: all manufacturer specific parameters / functions supported

Profile: profile defined parameters / functions supported only

Autom.: (default setting) PROFIBUS configuration (UFC400 or Profile) will be

selected automatically by the PROFIBUS master itself without interaction by

the user. The active selection is configured during startup by using the

Set_Prm and/or Set_Slave_Add service. No additional acyclic parameter

settings required.

C5.8.8

Condensed Status

Defines the general method how the device handles status and diagnostics:

On: Condensed status and diagnosis information according PROFIBUS

Profile 3.02 will be used

Off: Classic status and diagnosis information according PROFIBUS Profile

3.02 will be used

No.

Function

Settings / descriptions

D4.0

Service Parameter

D4.7.0

Identification No.

Same as C5.8.6

D4.8.0

PB Cold Start

Initiates a PROFIBUS cold start.

During a PROFIBUS cold start nearly all parameter values of the whole

device will be set to their default values (Exceptions are the PROFIBUS

address and the identification No.). The PROFIBUS cold start will be carried

out without disconnecting an already established connection to a PROFIBUS

master system.

Select:

No (terminates this function) / Yes (PROFIBUS cold start will be carried out

at once)