Lede no tare ledf no m.def ledg no dep.to, Led c no tor2 ledd no rs, Disp total fct1 fct2 fct3 – KROHNE C95 CI EN User Manual
Page 17: Fct4, Fct5, Hold clr.m tare
S
Led marking e
No dedication (led still)
Led lit when a tare has been entered
Led marking f
No dedication (led still)
Led lit if defect recorded
Led marking g
No dedication (led still)
Led lit if overstepping of the 1st totalizer
-99 999 > counting > 999 999
(Hi.tot
≠0)
LEdE
no
tArE
LEdF
no
M.dEF
LEdG
no
dEP.to
Led marking c
No dedication (led still)
Led lit when logic input 2 is active
Led marking d
No dedication (led still)
Led lit during answering of the instrument on
the RS
LEd C
no
tor2
LEdd
no
RS
Switching to the cumulated value display if LOGIC activated,
if not instant value display.
Function on the totalizer
meter zero reset on ascending front, and counting.
zero reset on ascending front, and counting on low level
only
diSP
totAL
Fct1
Fct2
Fct3
24VDC
0V
counting counting
counting
meter
meter
zero reset
zero reset
24VDC
0V
counting
counting stopped
counting
meter meter
zero reset
zero reset
zero reset on ascending front, and counting on high level
only
Counting on high level only (no reset)
Counting on low level only (no reset)
24VDC
0V
counting
counting stopped
meter
meter
zero reset
zero reset
24VDC
0V
counting
counting stopped
counting
counting stopped
Fct4
24VDC
0V
counting stopped
counting
Fct5
4.5.6 LOGIC inputs
(optional)
• Board of 2 LOGIC inputs : Input signal 24 Vdc
Possible functions :
Holding of the instant value display in case of activation of the
LOGIC function. The display and the analog output if it is dedi-
cated to the instant value remain fixed in case of variation of the input
signal. The relays and the totalizer carry on reacting to the input signal.
Min. and max. zero reset. The activation of the LOGIC
function resets the min. and max. to zero
Activation of function tare.
The meter switches to mode tare, the tare being the display
value present at the time of this activation.
HoLd
CLr.M
tArE