Menu "Calibrate"
Rake height
In case of a bottom rake or a slope, this parameter is subtracted
from the maximum depth to set the actual stop. At this point the
sensor will stop and return home in the case that there is no
sludge (refer to on page 24 for information concerning the
positions and zones).
Track
This option determines where the measuring cycle stops. The
setting "Level" effects that the measurement stops when the
sensor detects the sludge blanket (as defined in the submenu
"Blanket Cons"). The setting "Profile" runs each measuring cycle
over the complete measuring zone, i.e. the sensor goes to the
maximum immersion depth.
Submenu
Option
Description
Mode
Mode
Mode
Mode
There are two operation modes:
Setup
Allows to change the calibration parameters or to use the
"MANUAL UP/DOWN" button.
Trig
Normal working mode, in which the device will take a sample on
every external trig.
Manual
Manual
Manual
Manual
For further information see submenu "Manual" in the menu
"Settings".
Take sample
Take sample
Take sample
Take sample
Choice of the calibration type.
No
To abort calibration.
Zero
To perform a zero point calibration.
Sample
To perform a sludge calibration.
Cons
Cons
Cons
Cons
Makes the display showing only the current consistency.
Sample
Sample
Sample
Sample
This function provides the opportunity to enter the consistency
(i.e. suspended solids value) from a laboratory test.