OPERATION

32

OPTISENS OAM 2080

www.krohne.com

08/2010 - MA OPTISENS OAM 2080 R01 en

Menu "Calibrate"

Rake height

In case of a bottom rake or a slope, this parameter is subtracted

from the maximum depth to set the actual stop. At this point the

sensor will stop and return home in the case that there is no

sludge (refer to on page 24 for information concerning the

positions and zones).

Track

This option determines where the measuring cycle stops. The

setting "Level" effects that the measurement stops when the

sensor detects the sludge blanket (as defined in the submenu

"Blanket Cons"). The setting "Profile" runs each measuring cycle

over the complete measuring zone, i.e. the sensor goes to the

maximum immersion depth.

Submenu

Option

Description

Mode

Mode

Mode

Mode

There are two operation modes:

Setup

Allows to change the calibration parameters or to use the

"MANUAL UP/DOWN" button.

Trig

Normal working mode, in which the device will take a sample on

every external trig.

Manual

Manual

Manual

Manual

For further information see submenu "Manual" in the menu

"Settings".

Take sample

Take sample

Take sample

Take sample

Choice of the calibration type.

No

To abort calibration.

Zero

To perform a zero point calibration.

Sample

To perform a sludge calibration.

Cons

Cons

Cons

Cons

Makes the display showing only the current consistency.

Sample

Sample

Sample

Sample

This function provides the opportunity to enter the consistency

(i.e. suspended solids value) from a laboratory test.

Submenu

Option

Description