Step 4: Adjusting

• Open the door of the meter for 45°.
• Adjust the meter horizontally using the last two screws that you screw in the lower part of the

frame.

• Turn the screws until the door does not move any more in the 45° position.

3.4.2 Installation of flushing

CAUTION!
Always regard the following items when installing and using the flushing function:

Do not flush the device at temperatures below 0°C / 32°F as there is the risk of damage by
freezing water within the flush hoses and the solenoid valve!

Assure that all outdoor installations protect the hoses and the valve from freezing conditions.
If this is not possible, disconnect and empty all hoses!

Assure that the flushing water has a pressure between 2 bar / 29 psi and 6 bar / 87 psi. Below
the minimum pressure a sufficient cleaning is not possible, a water pressure above the
maximum value can cause damages.

Always use filtered water that is free of visible solids as they can cause damages. If this is the
case, you can also use town water.

INFORMATION!
The minimum diameter of the flush hose has to be at least 10 mm / 0.39".