OPERATION

5

27

OPTISENS OAM 2080

www.krohne.com

08/2010 - MA OPTISENS OAM 2080 R01 en

i

The device measures the consistency of the clean water or the sludge. In case of zero
calibration the displays shows the message "Zero calibration done" at the end, in case of
sludge calibration the display shows the measured value in a new pop-up window.

• In case of a sludge calibration press ^ to store the value and take the sample to a laboratory

to determine the sludge content.

• Irrespective of the calibration type press ^ to close all messages and return to the submenu

level of the main menu "Calibrate".

• If you want to return to the normal measuring mode, go to the submenu "Mode" again and

choose the option "Trig" (procedure is similar to the one on the beginning of this
actionsequence).

i

The green LED on the door now burns steadily.

• Press ↑ and ^ for 5 seconds to go to the normal measuring mode.

Entering the laboratory result (sludge calibration)

• Press ↑ and ^ simultaneously for 5 seconds to open the sensor menu.
• Use ↓ to go to the menu "Calibrate" and enter it by pressing ^.
• Use ↓ to choose the submenu "Sample" and enter it by pressing ^.
• Enter the laboratory result with the help of ↑ or ↓and confirm by pressing ^.

i

The converter will automatically correct the sludge consistency value from now on.

• Return to the normal measuring mode by pressing ↑ and ^ for 5 seconds.

5.6 Manual hauling

The meter has a manual hauling function. With the help of the safety switch you can override the
main board control of the cable drum motor and make the sensor go up or down with the manual
hauling switch ( refer to

Description of the single components on page 10). Forcing the sensor to

go up might be necessary for example if the meter stops functioning correctly during an active
sample.

For manual hauling execute the following steps and :

• Open the enclosure with the help of the key you find in the scope of delivery.
• Press the safety switch coming out of the front side of the connection terminal box to supply

the device with power.

• To make the sensor go up or down use the manual hauling switch on the top of the connection

terminal box (the switch has three positions: up/neutral/down).