Installation, 4 installation order – KROHNE OPTISENS OAM 2080 EN User Manual
Page 14
3
INSTALLATION
14
OPTISENS OAM 2080
www.krohne.com
08/2010 - MA OPTISENS OAM 2080 R01 en
3.4 Installation order
Steps to install the measuring system
• Mounting of converter
(for detailed information refer to the relevant chapter in the converter manual).
• Mounting of sludge blanket meter
(for detailed information refer to the relevant section in this chapter).
• Installation of flushing, if required
(for detailed information refer to the relevant section in this chapter).
• Connection of the rake guard switch, if required
(for detailed information refer to the relevant section in this chapter).
• Electrical installation of converter
(for detailed information refer to the relevant chapter in the converter manual)
• Electrical installation of the sludge blanket meter
(for detailed information refer to the relevant section in this chapter).
• Setup of converter
(for detailed information refer to the relevant chapter in the converter manual).
• Settings and calibration of the sludge blanket meter
(for detailed information refer to the relevant chapter in the manual of the meter).
INFORMATION!
The OAM 2080 only works in conjunction with the MAC 080 multiparameter converter which you
need for configuration, local display and transmission of the measuring results. Additionally the
converter must have the software version 3.1 or higher. The converter and the mounting plate
belong to the standard scope of delivery.
DANGER!
Do not install the sludge blanket meter in hazardous areas, it can ignite explosive gases!
INFORMATION!
To install the measuring system in the best way, follow the steps described below. Regard that
different steps may vary in meaning, depending on the particular sensor and the number of
sensors that you want to connect to the converter.
INFORMATION!
All MAC 080 converters with a program version of 3.1 or higher are able to detect and install the
sensor automatically. This procedure starts when both the converter and the sensor are
supplied with power and the electrical connection between the devices is established.
.book Page 14 Friday, August 20, 2010 1:07 PM