5.7 Display of sludge profile (optional)

When using the profile function, the single channels transmit the following measuring results
(refer to on page 24 for information concerning the positions and zones):

Execute the following steps to activate the profile function:

• Press ↑ and ^ simultaneously to enter the sensor menu.
• Use ↓ to go to the main menu "Settings" and enter it by pressing ^.
• Use ↓ to go to the submenu "Analog" and enter it by pressing ^.
• Use ↓ to choose "All" and confirm by pressing ^.
• Press ↑ until you reach the top and press ^ to exit the submenu "Analog".

i

You are in the main menu "Settings" again.

• Use ↓ to go to the submenu "Second" and enter it by pressing ^.
• Use ↓ to choose "Fluff" and confirm by pressing ^.
• Press ↑ until you reach the top and press ^ to exit the submenu "Second".

i

You are in the main menu "Settings" again.

• Use ↓ to go to the submenu "Advanced setup" and enter it by pressing ^.
• Use ↓ to go to the option "Track" and enter it by pressing ^.
• Use ↓ to go to "Profile" and confirm by pressing ^.
• Exit the submenu "Advanced setup" by pressing ↑ and ^ for five seconds.
• Leave the sensor menu by pressing ↑ and ^ for 5 seconds.

i

The meter is in the normal measuring mode now.

INFORMATION!
This function requires an expansion module with two more 4...20 mA outputs in the MAC 080
converter. Altogether the converter has four outputs with the extra card. For detailed
information about the installation of the converter expansion module refer to the section
"Description of additional analog outputs" in the converter manual.

Channel

Measuring result

1

Height of sludge blanket, every new measurement updates the signal. The

scale is defined by the maximum and minimum value entered in the menu

"Scale / Alarm".

2

Height of fluff blanket, every new measurement updates the signal.

3

Consistency, displayed in real-time during a measurement (otherwise 0 mA if

no measurement takes place), scale is locked to 0...10000 mg/l.

4

Depth of sensor, displayed in real-time during a measurement (otherwise the

output channel remains on 20 mA if no measurement takes place).