INSTALLATION

6

IFC 300

www.krohne.com

08/2010 - 4000070303 - QS IFC 300 R06 en

Possible scope of delivery for signal converter / measuring sensor

2.3 Storage

• Store the device in a dry, dust-free location.
• Avoid continuous direct sunlight.
• Store the device in its original packing.
• Storage temperature: -50...+70°C / -58...+158°F

2.4 Transport

Signal converter

• No special requirements.

Compact version

• Do not lift the device by the signal converter housing.
• Do not use lifting chains.
• To transport flange devices, use lifting straps. Wrap these around both process connections.

2.5 Installation specifications

Measuring sensor

Measuring sensor + signal converter IFC 300

Compact

Remote field
housing

Remote wall-
mounted housing

Remote rack-mounted
housing
R (28 TE) or (21 TE)

OPTIFLUX 1000

OPTIFLUX 1300 C

OPTIFLUX 1300 F

OPTIFLUX 1300 W

OPTIFLUX 1300 R

OPTIFLUX 2000

OPTIFLUX 2300 C

OPTIFLUX 2300 F

OPTIFLUX 2300 W

OPTIFLUX 2300 R

OPTIFLUX 4000

OPTIFLUX 4300 C

OPTIFLUX 4300 F

OPTIFLUX 4300 W

OPTIFLUX 4300 R

OPTIFLUX 5000

OPTIFLUX 5300 C

OPTIFLUX 5300 F

OPTIFLUX 5300 W

OPTIFLUX 5300 R

OPTIFLUX 6000

OPTIFLUX 6300 C

OPTIFLUX 6300 F

OPTIFLUX 6300 W

OPTIFLUX 6300 R

OPTIFLUX 7000

OPTIFLUX 7300 C

-

-

-

WATERFLUX 3000

WATERFLUX 3300 C

WATERFLUX 3300 F

WATERFLUX 3300 W WATERFLUX 3300 R

TIDALFLUX 4000

-

TIDALFLUX 4300 F

-

-

INFORMATION!
The following precautions must be taken to ensure reliable installation.

Make sure that there is adequate space to the sides.

Protect the signal converter from direct sunlight and install a sun shade if necessary.

Signal converters installed in control cabinets require adequate cooling, e.g. by fan or heat
exchanger.

Do not expose the signal converter to intense vibration. The flowmeters are tested for a
vibration level in accordance with IEC 68-2-3.