Page 32

background image

3

ELECTRICAL CONNECTIONS

32

IFC 300

www.krohne.com

08/2010 - 4000070303 - QS IFC 300 R06 en

3.5.2 Connecting the signal and field current cables, wall-mounted housing

• The outer shield of signal cable A and/or B is connected via the stranded drain wire.
• If a shielded field current cable is used, the shield must NOT

NOT

NOT

NOT be connected in the housing of

the signal converter.

• Bending radius: ≥ 50 mm / 2"

1 Open the housing cover.
2 Pass the prepared signal cable through the cable entry and connect the corresponding strand-

ed drain wires and conductors.

3 Connect the stranded drain wire of the outer shield.
4 Pass the prepared field current cable through the cable entry and connect the corresponding

conductor.
Any shield that is present must NOT

NOT

NOT

NOT be connected.

5 Tighten the screw connections of the cable entry and close the housing cover.

Figure 3-14: Electrical connection of the signal and field current cables, wall-mounted housing

INFORMATION!
Ensure that the housing gasket is properly fitted, clean and undamaged.