3.6.8 Preparing field current cable C, connection to signal converter

• Field current cable C is not part of the scope of delivery.
• Bending radius: ≥ 50 mm / 2"

Required materials:

• Shielded 2-wire copper cable, with fitted heat-shrinkable tubing
• DIN 46 228 wire end ferrules: size according to the cable being used

1 Strip the conductor to dimension a.
2 Trim the outer shield to dimension b and pull it over the outer sheath.
3 Crimp wire end ferrules onto both conductors.
4 Pull a shrinkable tube over the prepared cable.

DANGER!
A shielded two-wire copper cable is used as the field current cable. The shielding MUST

MUST

MUST

MUST be

connected in the housing of the measuring sensor and signal converter.

INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.

Figure 3-29: Preparation of field current cable C

a = 80 mm / 3.15"
b = 10 mm / 0.4"