Electrical connections – KROHNE IFC 300 Converter Quickstart EN User Manual
Page 19
ELECTRICAL CONNECTIONS
3
19
IFC 300
www.krohne.com
08/2010 - 4000070303 - QS IFC 300 R06 en
Wall mounted housing
• The connection of the outer shield (60) is carried out in the wall-mounted housing via the
stranded drain wire (6).
• Bending radius: ≥ 50 mm / 2"
Required materials
• Push-on connector 6.3 mm / 0.25", insulation to DIN 46245 for conductor Ø = 0.5...1 mm
2
/
AWG 20...17
• PVC insulation tubing, Ø2.5 mm / 0.1"
• Heat-shrinkable tubing
• Wire end ferrule to DIN 46 228: E 1.5-8 for the stranded drain wire (1)
• 2x wire end ferrules to DIN 46 228: E 0.5-8 for the insulated conductors (2, 3)
1 Strip the conductor to dimension a.
2 Cut off the inner shield (10) and the outer shield (60). Make sure not to damage the stranded
drain wires (1) and (6).
3 Slide the insulation tubing over the stranded drain wires.
4 Crimp the push-on connector onto the stranded drain wire (6).
5 Crimp the wire end ferrules onto the conductors (2, 3) and stranded drain wire (1).
6 Pull the heat-shrinkable tubing over the prepared signal cable.
INFORMATION!
Assembly materials and tools are not part of the delivery. Use the assembly materials and tools
in compliance with the applicable occupational health and safety directives.
Figure 3-3: Signal cable A, preparation for wall-mounted housing
a = 80 mm / 3.15"