beautypg.com

Electrical connections, 3 length of signal cable a – KROHNE IFC 300 Converter Quickstart EN User Manual

Page 20

background image

3

ELECTRICAL CONNECTIONS

20

IFC 300

www.krohne.com

08/2010 - 4000070303 - QS IFC 300 R06 en

3.4.3 Length of signal cable A

INFORMATION!
For temperatures of the medium above 150

°

C / 300

°

F, a special signal cable and a ZD

intermediate socket are necessary. These are available including the changed electrical
connection diagrams.

Measuring sensor

Nominal size

Min. electrical
conductivity
[µS/cm]

Curve for signal
cable A

DN [mm]

[inch]

OPTIFLUX 1000 F

10...150

3/8...6

5

A1

OPTIFLUX 2000 F

25...150

1...6

20

A1

200...2000

8...80

20

A2

OPTIFLUX 4000 F

2.5...150

1/10...6

1

A1

200...2000

8...80

1

A2

OPTIFLUX 5000 F

2.5...100

1/10...4

1

A1

150...250

6...10

1

A2

OPTIFLUX 6000 F

2.5...150

1/10...6

1

A1

WATERFLUX 3000 F

25...600

1...24

20

A1

Figure 3-4: Maximum length of signal cable A

1 Maximum length of signal cable A between the measuring sensor and signal converter [m]
2 Maximum length of signal cable A between the measuring sensor and signal converter [ft]