B.31 [source:]current[:level]:protect:mode command, B.32 [source:]current[:level]:protect:mode? query, B.31 – KEPCO BOP 1KW-MG Operator Manual, Firmware Ver.4.08 to 4.11 User Manual

Page 165: Sour, B.32, B.33, B.34, B.35, B.36, B.37

background image

BOP-1K 070212

B-11

B.31 [SOURce:]CURRent[:LEVel]:PROTect:MODE COMMAND

CURR:PROT:MODE

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:MODE (EXT | FIX | LESS |)
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:MODE (EXTernal | FIXed | LESSer |)

Description: Determines how current protection limits are controlled.

FIXED - Allows the limits to be controlled by digital signals from either the keypad (if the unit is in local
mode) or SCPI command (if the unit is in remote mode).
EXTERNAL: Allows the limits to be controlled by analog signals applied to the analog port; limits set
from the keypad or SCPI command have no effect (see PAR. 3.3.6.1).
LESSER LIMIT: Allows protect limit to be automatically selected from either 1) the external analog limit
CURRent applied to the Analog I/O port or 2) the value set from the keypad or SCPI command).
Whichever limit has a lower absolute value (closest to zero) has effect

B.32 [SOURce:]CURRent[:LEVel]:PROTect:MODE? QUERY

CURR:PROT:MODE?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:MODE?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:MODE?
Returns FIXED, EXTERNAL or LESS

Description: Identifies how current protection limits are controlled. (See PAR. B.31 for details).

B.33 [SOURce:]CURRent[:LEVel]:PROTect:NEGative COMMAND

CURR:PROT:NEG

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:NEG
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:NEGative
where is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
current.

Description: Establishes the negative protection limit for current, maximum is 1% greater than rated (nominal) out-

put current.

B.34 [SOURce:]CURRent[:LEVel]:PROTect:NEGative? QUERY

CURR:PROT:NEG?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:NEG?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:NEGative?
Returns

Description: Identifies the negative protection limit for current.

B.35 [SOURce:]CURRent[:LEVel]:PROTect:POSitive COMMAND

CURR:PROT:POS

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:POS
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:POSitive
where is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
current.

Description: Establishes the positive protection limit for current, maximum is 1% greater than rated output current.

B.36 [SOURce:]CURRent[:LEVel]:PROTect:POSitive? QUERY

CURR:PROT:POS?

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:POS?
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:POSitive?
Returns

Description: Identifies the positive protection limit for current.

B.37 [SOURce:]CURRent[:LEVel]:PROTect:LIMit[:BOTH] COMMAND

CURR:PROT:LIM[:BOTH]

Syntax:

Short Form: [SOUR:]CURR[:LEV]:PROT:LIM[:BOTH]
Long Form: [SOURce:]CURRent[:LEVel]:PROTect:LIMit:[BOTH
where is between minimum (box) value (Figure 1-3) and 1% above rated (nominal) output
current.

Description: Establishes the maximum value possible for both the positive and negative protection limits for cur-

rent. Requires MEM:UPD (PAR. B.12) to save for power up. See PAR. 3.5.2.2 and Figure 3-16 for
special programming considerations.