9. Battery Check
(1)First Alarm
When the battery voltage becomes less than 4.5 V, the message
"BATTERY" will be indicated at the top of the display (First Alarm). In approx.
5 minutes after the First Alarm, the display will switch to the following screen
(Second Alarm). When the Second Alarm is given, the pump, laser radiation
and software will stop, and the POWER key will become ineffective.
If the battery level becomes low during a measurement, the battery must be
replaced with an AC adapter. Power supply will automatically switch to AC
power supply when the AC adapter is inserted in the power inlet.
For continuous measurements for long periods, please use the AC adapter.
■
Please refer to the following table for the data storage conditions
when battery alarm is indicated.
While the First Alarm is displayed, data storage is possible.
Measurement
Mode
Data Storage
REPEAT
Every data measured before the Second Alarm will be
stored
SINGLE
If the measurement is finished before the Second
Alarm, the data will be stored.
CONTINUOUS
Data will be stored if the "stop" key is pressed during
the First Alarm.
(2)Second Alarm
CALCULATION
The calculation result which is provided based on the
data measured before the Second Alarm will be stored.
BATTERY
REPEAT 003BPM 15:25
WAIT 0.3 0.00E+0/cf
01/06 0.5 0.00E+0/cf
25.5℃ 56.0% 0.25m/s
BATTERY