9. Battery Check









(1)First Alarm

When the battery voltage becomes less than 4.5 V, the message
"BATTERY" will be indicated at the top of the display (First Alarm). In approx.
5 minutes after the First Alarm, the display will switch to the following screen
(Second Alarm). When the Second Alarm is given, the pump, laser radiation
and software will stop, and the POWER key will become ineffective.
If the battery level becomes low during a measurement, the battery must be
replaced with an AC adapter. Power supply will automatically switch to AC
power supply when the AC adapter is inserted in the power inlet.

For continuous measurements for long periods, please use the AC adapter.

Please refer to the following table for the data storage conditions
when battery alarm is indicated.

While the First Alarm is displayed, data storage is possible.

Measurement

Mode

Data Storage

REPEAT

Every data measured before the Second Alarm will be

stored

SINGLE

If the measurement is finished before the Second

Alarm, the data will be stored.

CONTINUOUS

Data will be stored if the "stop" key is pressed during

the First Alarm.






(2)Second Alarm

CALCULATION

The calculation result which is provided based on the

data measured before the Second Alarm will be stored.





BATTERY

REPEAT 003BPM 15:25
WAIT 0.3 0.00E+0/cf
01/06 0.5 0.00E+0/cf
25.5℃ 56.0% 0.25m/s

BATTERY