KANOMAX 3886 Handheld Particle Counter User Manual

Page 20

background image

13


4.3 Setting of Alarm level



When PREV key is pushed at screen, the screen will switch to
the

screen.









In case the particle concentration exceed the acceptable level, or
temperature,
Relative humidity or air velocity go out of the specified range, this unit
can notify the occurrence of these alarm situation.
This menu is to preset the alarm level.


If SET key is pushed when the cursor is here, it moves to the following
page.

△ ▽ key

Adjust the value

SET key

Move the cursor





PREV key

Go back to MENU after the setup


Lower bound Upper bound

unit

Setting range

1 SMALL Small

particle −

0∼7.00E+7

2 LARGE Large

particle −

0∼7.00E+7

3 TEMP Temperature ○

0∼122.0

4 HUMI Humidity

%RH

0∼100.0

5 VEL Air

velocity ○

0∼200.0

*:

Selected unit (refer to 4.4)


To activate the alarm buzzer, change the “BEEP: N” to “BEEP: Y” on the setup screen of the measurement mode.
When the sampling time has expired, the measured value will be compared with the setting range, and the
measured value will blink if it is out of the setting range.
Once you have an alarm condition, the alarm will not be reset unit the data of the following measurement falls in
the setting range.
When the unit of particle data is set to “COUNT”, the reading will blink at the time the measured value exceeds the
setting range.
To stop the buzzer, push any key except the POWER key.

1. SMALL 0.00E+0

2. LARGE 0.00E+0

3. TEMP 12.5-30.0

<

ALARM>

4. HUMI 30.0-60.0
5. VEL 0.10-0.85

<

MENU>

1. MODE 4.DATA

2. ALARM 5.UTILITY

3. UNIT


1.

REPEAT 4.CALC

2.SINGLE 5.REMOTE
3.CONT

MODE

PREV

1.MODE ・・・・・・ Selection of measurement mode and setup of

parameter

2.ALARM ・・・・・ Setup of alarm level
3. UNIT ・・・・・・・・・ Selection of optional probe and unit
4. DATA ・・・・・・・・ Request for stored data in built-in memory
5. UTILITY ・・・・・・ Calendar and computer communication setting

△、▽

SET