KANOMAX 3886 Handheld Particle Counter User Manual
Page 29

22
6.3 Dump of stored data… <4.DATA>→<2.DUMP>
Put the optional RS-232C cable into the modular jack of GEO-α,
and connect the other end of the cable with the PC to transfer
the stored data.
Do the communication setting of the PC and make the
condition that the PC can readout the data.
1
START
The first tag number of the stored data
to be transmitted
2
END
The last tag number of the stored data
to be transmitted
△ ▽ key
Adjust the value
SET key
Move the cursor
START/STOP key
Start transmitting of the stored data
PREV key
Go back to DATA
Required Items
- Computer
- Application software, Model S388-61 (Optional): Measurement software
- RS-232C cable, Model 3886-08 (Optional): Communication cable for connecting GEO-α and PC
Setting up the computer
Function
GEO-α Setting
Word length
8 bit
Parity bit
None
Set parity
Odd number
Baud rate
9600
Signal cable
GEO-α
Connection
Computer (D-sub 9 pin)
Pin number
Signal name
Pin number
Signal name
1 TXD
2 RXD
3 RXD
3 TXD
5 CTS
7 RTS
6 GND
5 GND
4 DTR
6 DSR
Modular jack
STORE 139
1. START 001
2. END 139
Start transmitting