KANOMAX 3886 Handheld Particle Counter User Manual
Page 26

19
5.5 CALCULATION Mode
It is a mode by which measures repeatedly, and mean value from the measurement data, a standard deviation, the maximum
value, and minimum value are calculated. Only result is preserved, and each measurement result is not preserved in the data
store. The measurement frequency can be set at the grain size, the data store, warning, the printer, and the sample time.
Display screen
Operation
key
Operation explanation
POWER
Push POWER key to input the power supply.
PREV
Push PREV key twice to proceed MODE screen
△ ▽ SET Select 4. CALC
△ ▽ SET
■
Setup the particle size, requirement of data
storage , alarm , printout, sampling time,
frequency.
Use △ ▽ key to change the setting
Condition, then push SET key.
After the input is done, change NO to OK
and push SET key.
WAIT Mode
WAIT sign is shown for the stabilization of internal pump
READY Mode
WAIT sign turns to READY in 10 seconds and
measurement can be started.
Push START/STOP key to start the measurement
START/STOP
The display shows the real-time data.
Remaining time of each measurements
⑧:
The sample time of the remainder is
displayed.
⑨:
Present measurement frequency
The last measurement data is maintained on the screen
for five seconds though the following measurement
begins at the same time as ending measuring time.
Display of
result.
SET
After the last data is displayed for five seconds, result is
displayed when the set measurement frequency ends.
The data of the small particle is displayed first .
It is possible to switch with the data of the
large
★
Data changes in order saying the
temperature, humidity, Air velocity , the
small particle, and the large particle
whenever the SET key is pushed when the
temperature humidity and Air velocity probe
are used.
PREV
It returns to measuring the set screen with the
PREV key.
★
When the measurement ends when setting the
printer is Y, the result is printed at once .(refer to 6.4)
★
When the START STOP key is pushed while
measuring, the measurement is stopped,
and the measurement data of times ahead of
that is used and operated.
CALC 003BPM 15:25
WAIT 0.3 0.00E+0/cf
01/06 0.5 0.00E+0/cf
CALC 003BPM 15:25
READY 0.3 0.00E+0/cf
01/06 0.5 0.00E+0/cf
CALC 004BPM 15:25
09:59 0.3 0.00E+0/cf
01/06 0.5 0.00E+0/cf
⑧
→
⑨
→
CALC 0.3/0.5µm
STR:Y BEEP:N PR:N
SAMPLE 10:00 06TIMES
OK
CALC AVE 0.00E+/cf
0.3µm S.D 0.00E+0/cf
06T MAX 0.00E+0/cf
MIN 0.00E+0/cf
CALC 0.3/0.5µm
STR:Y BEEP:N PR:N
SAMPLE 10:00 06TIMES
OK
1. REPEAT 4.CALC
2. SINGLE 5.REMOTE
3. CONT
CALC AVE 0.00E+/cf
0.5µm S.D 0.00E+/cf
0.6T MAX 0.00E+0/cf
MIX 0.00E+0 / cf