Handling & cautions – KANOMAX 3886 Handheld Particle Counter User Manual

Page 11

background image

4

3. Handling & Cautions


3.1 Power supply

Please use the supplied AC adapter and refrain from the battery operation for the long consecutive
measurements (more than 2 hours)
This instrument has the monitoring function of operating voltage, and battery alarm will be indicated
when the voltage goes down below the specified value.
If you leave the instrument in such a conditions for a few minutes, the power automatically goes off. In
some of measuring mode, the data of measurement in process will not be stored. (Please refer to
Chapter 8 for details)
If the alarm sign is shown, please stop the measurement and charge the batteries, or replace with the
charged batteries.


◆Use of AC Adapter

Insert the plug of the supplied AC adapter into the power inlet at the side of instrument.

The AC power should be in the range of 86-264V 50/60Hz. Do not use the AC power outside of this
range.

◆Use of Ni-MH Batteries

Prepare 4 pieces of Ni-MH batteries (1.2V, 2500mAh) and fully charge them. Charging time is approx.

260 minutes.

When charging is completed, put the batteries into the instrument in the correct directions.

Battery life is about 3 hours, but it will vary by the type and capacity of battery, or status of

charging. When optional Temp.&Humidity probe and Air velocity probe are used at the same

time, there will be the cases that operating hours will become less than 2 hours.











Figure: Direction of Inserting Batteries


Though alkaline batteries can also be used, please note that the battery life for alkaline batteries will be
approx. 1.5 hours which is shorter than the Ni-MH batteries.