6.4 Printout of stored data… <4.DATA>→<3.PRINT>



Put the optional Printer cable into the modular jack of GEO--α,
and connect other side of the cable with the Printer to print
the stored data.

1

START

The first tag number of the stored data
to be printed

2

END

The last tag number of the stored data
to be printed

△ ▽ key

Adjust the value

SET key

Move the cursor

START/STOP key

Start printing of the stored data

PREV key

Go back to DATA

Preparations

Printer (Option) ・・・・・・・・ Recommendatory Printer Model. DPU-201GS (SEIKO CO., LTD),
Signal cable (Model 3886-07: Option), Signal cable connect GEO-α with Printer

DIP Switch Setting

Switch Number

Function

GEO-α Setting

Printer

SW1

Word length

8 bit

ON

SW2

Parity bit

None

ON

SW3

Set parity

Odd number

ON

SW4∼6

Baud rate

9600

following table

Baud rate

SW4 Setting

SW5 Setting

SW6 Setting

9600 OFF OFF ON

* When using DPU-H245, please use it with the manufacturers default setting.

Signal cable

GEO-α Printer

Pin number

Signal name

Pin number

Signal name

1 TXD 3 DATA
6 GND 4 GND
5 CTS 8 BUSY
6 GND 5 GND

<>

When the measurement interval (INT) in REPEAT mode is set at 15 seconds or less, there is a possibility

of Buffer error when printing is executed during a measurement. To print during a measurement, please
set the measurement interval over 15 seconds.

Modular jack

Start printing

STORE 139

1. START 001
2. END 139