l
The cable does not slide or move after the connectors are tightened.
l
Sufficient cable has been threaded through the connector so that secure wiring
connections can be made.
l
There is no weight or force from the water pulling on the cable.
Ensuring Good Electrical Connections
To ensure that all sensors and power sources function properly consider the following
recommendations.
l
Make sure that each wire is stripped-and-tinned to .025 in.
l
Make sure that each wire is tightly screwed into the terminal strip.
l
Make sure that each wire is touching the terminal strip. If the plastic wire jacket is
clamped into the terminal strip, connections will not be made.
l
Clip or cap any unused wires.
l
Make sure that ribbon cables and the coin cell battery are secure.
AC Electrical Connections
Danger—Make sure that power to the controller is disconnected before making
any wiring connections.
Connexions Aux Courant Secteur/Alternatif
Danger—Débrancher toute alimentation à l'appareil avant de connecter les fils.
Pin
Description
North American Wire Color
European Wire Color
L
Line
Black
Brown
N
Neutral
White
Blue
Protective Earth
Green
Green and yellow
AC Wiring Description
Figure 4.1 Protective Earth ground screw location and connection