beautypg.com

4 electrical connections, Power options, Electro-static discharge (esd) recommendations – In-Situ Con TROLL PRO Installation Manual (purchased after 10 May 2012) User Manual

Page 16: Customer-supplied electrical equipment, Ensuring dome connectors are water-tight, Electrical connections

background image

16

Electrical Connections

Danger—Only properly trained and qualified personnel should install the

controller described in this manual. This instrument should be installed for use in non-
hazardous locations only.

Danger—L'installation de l'appareil “controller” décrite dans le guide d'utilisateur

doit impérativement être réalisée par des personnes qualifiées. L'installation de cet
appareil est seulement prescrite pour des emplacements sans risques.

Power Options

The Con TROLL PROis available with 2 power options.

l

DC-powered, which requires a 9-36 VDC supply as the primary power source.
This model can trigger low-voltage (<50 V) relays only. See "Input/Output
Connections" on page 20.

l

AC-powered, which requires a 100-240 VAC as the primary power source. This
model can trigger low-voltage relays (<50 V) through the I/O board only (See
"Input/Output Connections" on page 20.), and high-voltage relays (>50 V) thro
ugh
the AC board only.

Electro-Static Discharge (ESD) Recommendations

Before making wiring connections, discharge any static electricity from your body

before touching circuit boards and other internal components by touching a grounded
metal object.

When making wiring connections, make sure to remain properly grounded by

wearing an ESD wrist strap or similar device.

Customer-Supplied Electrical Equipment

The end user must supply the following items.

l

18-12 AWG wire for electrical connections using conduit or 100-240 VAC power
supply

l

24-12 AWG cable for connecting relays and PLC devices

l

For hard-wired locations using conduit, a 100-240 VAC or 9-36 VDC source with
over current/disconnect protection will be provided by the end user.

l

For cord-connected plug and socket locations, approved suitable wiring to be
provided by the end user.

Ensuring Dome Connectors are Water-Tight

The dome connectors that you installed must be tightened securely after sensor cable
or electrical cables are routed through them.

These dome connectors are not meant to be weight-bearing.

For best results, ensure that the following conditions are met.

This manual is related to the following products: