beautypg.com

HT instruments М75 User Manual

Page 44

background image

M72 - M73 - M74 - M75

EN - 42

6.1. REFERENCE

GUIDELINES

Safety: IEC/EN61010-1,

IEC/EN61557-1-2-3-4-6-7

Insulation:

double insulation

Pollution level:

2

Measurement category:

CAT III 550V (phase to earth, phase to phase)

Maximum height of use:

2000m ; (6562ft)

LAN test

TIA568B

6.2. GENERAL

SPECIFICATIONS

Electrical features
Conversion:

ADC 16 bit, TRMS – True Root Mean Square

Measuring rate:

64 times per second

Display refreshing rate:

2 times per second

Mechanical features
Dimensions (L x W x H):

240 x 100 x 45mm ; (9 x 4 x 2in)

Weight (batteries included):

630g ; (22ounces)

Power supply
Battery type:

4x1.5V batteries type AA LR6 MN 1500

Low battery indication:

symbol "

" is displayed

Battery life:

Multimeter:

About 90 hours

:

> 1000 tests

LAN:

> 1000 tests

 0.2A:

> 1000 tests @ 1

M

:

> 1000 tests @ 480k

 (500VDC)

RCD:

> 1000 tests

Ra

:

> 1000 tests

AUTO:

> 1000 tests

Display
Features:

4 LCD with max. reading 9999 counts + symbol
and decimal point

6.3. ENVIRONMENT
6.3.1. Environmental

conditions

Reference temperature:

23°C ± 5°C ; (73°F ± 41°F)

Working temperature:

0°C ÷ 40 °C ; (32°F ÷ 104°F)

Relative humidity allowed:

<70%HR

Storage temperature:

-10 ÷ 60 °C ; (14°F ÷ 140°F)

Storage humidity:

<70%HR

This instrument complies with the requirements of the European Low Voltage

Directive 2006/95/CE (LVD) and EMC Directive 2004/108/CE

6.4. ACCESSORIES
Please see enclosed packing list.

This manual is related to the following products: