beautypg.com

Caution – HT instruments М75 User Manual

Page 36

background image

M72 - M73 - M74 - M75

EN - 34

7. Keep pressed GO for at least one second, the instrument performs the measurement

CAUTION

The message “Measuring” on the display means that the instrument is
measuring. During this phase never disconnect test leads.

8. At the end of the test, if the earth

resistance value is lower than 1999

,

the instrument emits a double sound
and displays a screen like this reporting
the value of the measured global earth
resistance and the current at which the
measurement has been performed

Measured value of
global earth
resistance

The symbol HOLD
remains on until a
new measurement
can be taken

Current used during
the measurement

9. If a test current of 100mA has been

selected and the earth resistance value
is lower than 1999

, by pressing FUNC

you will display alternatively the values
of the global earth resistance and the
prospective short circuit current phase
to earth, as well as the current at which
the measurement has been performed

Measured value of
global earth
resistance

The symbol HOLD
remains on until a
new measurement
can be taken

Current used during
the measurement

10. At the end of the test, if the global earth

resistance value is higher than 1999

,

the instrument emits a double sound
and displays a screen like this

Measured value of
global earth
resistance higher
than full scale

The symbol HOLD
remains on until a
new measurement
can be taken

CAUTION

To guarantee the correctness of measurements, a certain interval of time is
necessary between a measurement and the following. During this period the
symbol HOLD is displayed and no measurement can be taken. When the
symbol HOLD disappears, the instrument is ready for a new measurement.