4.9.

: RESISTANCE MEASUREMENT AND CONTINUITY TEST

CAUTION

Before taking resistance measurements make sure that the circuit under test is
not powered and that eventual condensers are discharged.

Fig. 5: Connection of the instrument’s terminals

1. Turn on the instrument

2. Press the arrow keys to select

3. If the measuring cables being used have not been calibrated, first calibrate them as

described in § 4.9.1

4. Insert the black and green cables in the

corresponding input terminals of the
instrument

Resistance value

5. Position the test probes on the desired

points of the circuit under test (see Fig. 5)

6. Example of display of resistance value. If

such value is lower than 40

 the

instrument emits an acoustic signal

7. To block the detected value on the display press FUNC/HOLD for less than 1 second

(see § 4.6.1)

8. To detect maximum, minimum and average values press FUNC/HOLD for more than 1

second (see § 4.6.2)

9. The measured value is out of accuracy if an input voltage is present

4.9.1. "CAL"

mode

1. Any addition or replacement of cables, extensions and croco clips nullify the previous

calibration and make necessary a new calibration before performing further
measurements. Therefore the instrument must be calibrated in the same conditions at
which it will operate during measurements

2. Short-circuit the cable ends with each other as shown in Fig. 6 making sure that the

metallic parts of test probes and crocodiles are in good touch