HT instruments М75 User Manual

Page 22

background image

M72 - M73 - M74 - M75

EN - 20

4.10.1. Anomalous cases which may occur during tests

1. If you wait more than 10 seconds

between the first measurement and the
second one, the instrument emits a
prolonged sound to signal the negative
outcome of the test and displays a screen
like this. It’s necessary to repeat the
entire procedure. Press GO and re-start
from point 6

2. If mode 1W is selected and the

instrument detects the connection of the
second probe like in mode 2W, a screen
like this is displayed to signal the error. A
prolonged sound is emitted until the error
condition is removed

3. If mode 2W is selected and the

instrument detects an input voltage
(between the two bushes) higher than
605V, a screen like this is displayed and
a prolonged sound is emitted until the
error condition is removed. Disconnect
the instrument promptly