beautypg.com

HT instruments М75 User Manual

Page 13

background image

M72 - M73 - M74 - M75

EN - 11

4.7.1. Anomalous

cases

which may occur during V Hz measurements

1.

The maximum input voltage is
550+10%V.
If the detected voltage value exceeds
605V TRMS
the instrument displays the
screen beside.
Disconnect immediately the instrument
from the circuit under test to avoid
electrical shocks and damages to the
instrument

2. If during a voltage measurement the

detected frequency value exceeds 400Hz
the instrument displays the screen beside

3. If during a frequency measurement the

detected value exceeds 400Hz the
instrument displays the screen beside

4. If during a frequency measurement the

detected value does not reach 30.0Hz
the instrument displays the screen beside










This manual is related to the following products: