beautypg.com

HT instruments М75 User Manual

Page 27

background image

M72 - M73 - M74 - M75

EN - 25

9. The continuity test is performed by supplying a current higher than 200mA if the

resistance value is lower than 5

 (including the cable resistance stored as offset after

calibration). For higher resistance values the instrument performs the test with
decreasing current

10. At the end of the test, if it has been

possible to generate at least 200mA
(not particularly high resistance
value), the instrument emits a
double sound to signal the positive
outcome of the test. The screen
beside is displayed

Resistance value

Test current value

11. At the end of the test, if it has not

been possible to generate 200mA
due to the high resistance value, the
instrument emits a prolonged sound
to signal the negative outcome of
the test. The screen beside is
displayed

Resistance value

Test current value


4.12.1. "CAL" mode

Fig. 10: Connection of the instrument’s terminals during calibration procedure

This manual is related to the following products: