Technical specifications – HT instruments М75 User Manual
Page 42
M72 - M73 - M74 - M75
EN - 40
6. TECHNICAL SPECIFICATIONS
Accuracy is indicated as [%reading + (digit number*resolution)] at: 23°C±5°C, <70%HR.
DC/AC TRMS Voltage
Range Resolution
Accuracy
DC
Accuracy
(30
70Hz)
Accuracy
(70
400Hz)
Input
impedance
1.0
999.9mV
0.1mV
(0.5rdg+2dgt) (1.0rdg+2dgt) (2.0rdg+2dgt) 1M
1.000
9.999V
0.001V
10.00
99.99V
0.01V
100.0
605.0V
0.1V
MAX, MIN, AVG, PEAK, resolution:
(5.0% rdg + 10 dgt); responce time: 500ms (MAX, MIN, AVG),1ms (PEAK)
Max crest factor: 3.0: V<1.0V; 1.5: V≥1.0V
DC/AC TRMS Current (with external jaws)
Range Resolution
Accuracy
DC
Accuracy
(30
70Hz)
Accuracy
(70
400Hz)
Input
impedance
Overload
protection
1.0
999.9mV
0.1mV
(0.5rdg+
2dgt)
(1.0 rdg +
2 dgt)
(2.0 rdg +
2 dgt)
1M
605V AC
max RMS
1.000
1.200V
0.001V
NOTE The mentioned accuracy does not consider the transducer's accuracy. Please refer to its user's manual
MAX, MIN, AVG, PEAK, resolution:
(5.0% rdg + 10 dgt); responce time: 500ms (MAX, MIN, AVG),1ms (PEAK)
Minimum input current: 1mV x transduction ratio of the clamp
Max crest factor: 3.0: V<1.0V; 1.5: V≥1.0V
Frequency measurement with test leads
Range Resolution
Accuracy
Input
impedance
30.0
199.9Hz
0.1Hz
(0.5rdg + 2dgt)
1M
200
400Hz
1Hz
Input voltage: 1mV
605.0V
Frequency measurement with jaws
Range Resolution
Accuracy
Overload
protection
30.0
199.9Hz
0.1Hz
(0.5rdg + 2dgt)
605V AC max RMS
200
400Hz
1Hz
Input voltage: 1mV
1V
Resistance and continuity test
Range Resolution
Accuracy
Buzzer Overload
protection
0.00
39.99 0.01
(1.0%rdg + 5dgt)
R<40
605V AC max RMS for 1 minute
40.0
399.9 0.1
400
3999
1
4.00
39.99 k 10
Phase sequence and phase conformity
Measuring method
Working voltage (V)
System
1 test lead (1W)
90
315 (Phase - Ground)
up to 315 V (Phase - Ground)
up to 550V (Phase - Phase)
2 test leads (2W)
90
315 (Phase - Neutral)
up to 315 V (Phase - Neutral)
up to 550V (Phase - Phase) (*)
Max crest factor: 1.5 ; Frequency range: 45
65 Hz
(*) The two-wire measurement can be performed also phase to phase in plants without neutral, even with one phase to earth, but
always with phase to phase voltage up to 550V
0.2A: Continuity with 200mA (M72, M74, M75)
Range Resolution
Accuracy
Overload
protection
0.00
19.99 0.01
(5.0% rdg + 3 dgt)
605V max RMS
20.0
99.9
0.1
Test current: >200mA DC up to 5
(measuring cables resistance included)
Accuracy of current measurement: 1mA
Open circuit voltage: 4 < V
0
< 24V