beautypg.com

1 precautions and safety measures, Caution – HT instruments HT7052 User Manual

Page 4

background image

HT7052

EN - 2

1 PRECAUTIONS

AND

SAFETY MEASURES

The instrument has been designed in compliance with standards IEC/EN61557-1 and
IEC/EN61010-1 regarding electronic measuring instruments

CAUTION

For the operator’s safety and to prevent damaging the instrument, follow the
procedures described in this manual and carefully read all notes preceded by
the symbol

Before and during measurements, carefully observe the following instructions:
 Do not perform any measurement in humid environments, in the presence of gas or

explosive or inflammable material or in dusty areas

 Even when no measurements are being performed, avoid any contact with the circuit

being tested, with exposed metal parts, with unused measuring leads or circuits, etc

 Do not perform any measurement when anomalies are found in the instrument, such as

deformations, breaks, substance leaks, no display view, etc

 Pay special attention when measuring voltages above 25V in special environments

(building yards, swimming pools, etc.) and 50V in ordinary environments, as there is the
danger of electric shocks

In this manual and on the instrument, the following symbols are used:

WARNING: Observe the instructions reported in the manual. An improper use
could damage the instrument and lead to dangerous situations for the operator

DC voltage or current

Dangerous voltages: risk of electric shocks

Instrument with double insulation

1.1 PRELIMINARY

INSTRUCTIONS

 This instrument has been designed for use in an environment with pollution level 2

 It may also be used to test industrial electrical systems up to CAT IV 600V to earth with

maximum voltage 600V between inputs

 Follow the usual safety rules to protect the operator from dangerous currents and

protect the instrument against improper use

 Never use the instrument resting on the floor, it must be placed over flat horizontal

surfaces

 Only the accessories supplied with the instrument guarantee safety standards. They

must be in good conditions and replaced, if necessary, with identical models

 Do not measure systems exceeding the current and voltage limit values specified

 Do not perform measurements in environmental conditions not within the limit values

indicated in this manual

 Before connecting the probes to the circuit to be tested, check that the correct function

is selected