1 PRECAUTIONS
AND
SAFETY MEASURES
The instrument has been designed in compliance with standards IEC/EN61557-1 and
IEC/EN61010-1 regarding electronic measuring instruments
CAUTION
For the operator’s safety and to prevent damaging the instrument, follow the
procedures described in this manual and carefully read all notes preceded by
the symbol
Before and during measurements, carefully observe the following instructions:
Do not perform any measurement in humid environments, in the presence of gas or
explosive or inflammable material or in dusty areas
Even when no measurements are being performed, avoid any contact with the circuit
being tested, with exposed metal parts, with unused measuring leads or circuits, etc
Do not perform any measurement when anomalies are found in the instrument, such as
deformations, breaks, substance leaks, no display view, etc
Pay special attention when measuring voltages above 25V in special environments
(building yards, swimming pools, etc.) and 50V in ordinary environments, as there is the
danger of electric shocks
In this manual and on the instrument, the following symbols are used:
WARNING: Observe the instructions reported in the manual. An improper use
could damage the instrument and lead to dangerous situations for the operator
DC voltage or current
Dangerous voltages: risk of electric shocks
Instrument with double insulation
1.1 PRELIMINARY
INSTRUCTIONS
This instrument has been designed for use in an environment with pollution level 2
It may also be used to test industrial electrical systems up to CAT IV 600V to earth with
maximum voltage 600V between inputs
Follow the usual safety rules to protect the operator from dangerous currents and
protect the instrument against improper use
Never use the instrument resting on the floor, it must be placed over flat horizontal
surfaces
Only the accessories supplied with the instrument guarantee safety standards. They
must be in good conditions and replaced, if necessary, with identical models
Do not measure systems exceeding the current and voltage limit values specified
Do not perform measurements in environmental conditions not within the limit values
indicated in this manual
Before connecting the probes to the circuit to be tested, check that the correct function
is selected