Caution – HT instruments HT7052 User Manual

Page 11

background image

HT7052

EN - 9

Auto-calibration prevents the reduction in accuracy when measuring very low currents. It
compensates the effects caused by ageing, temperature and humidity changes etc. A new
auto-calibration is recommended when the temperature changes by more than 5

C. If the

instrument detects an incorrect state during the autocalibration, the following warning
message will be displayed:

CAUTION

 TEST LEADS CONNECTED: DISCONNECT AND SWITCH ON THE

INSTRUMENT AGAIN

 CONDITIONS OUT OF RANGE: PRESS START TO CONTINUE


Possible reasons for out of range conditions are excessive humidity, excessively high
temperature, etc. In this case it is possible to perform measurements by pressing the
START/STOP button again but results could be out of technical specification

5.2 CONFIGURATION AND SETUP OF SYSTEMS PARAMETERS
The configuration and setup function enables the selection and adjustment of the
parameters (see Table 1) that are not directly involved in the measurement procedure (see
Fig. 5 and Fig. 6). In the lower section of the display the power supply status is shown

Fig. 5: Configuration menu

Fig. 6: Setup menu

PARAMETERS VALUE

DESCRIPTION

Memory clear

Clear all memory locations

Filter

Fil1, Fil2, Fil3, Fil0

Selection of noise rejecting filter (see § 6.3)

DIAG. Starting
time

0%…90%

Adjustment of start of the timer in the

DIAGNOSTIC TEST functions, according to the

nominal voltage Unominal. See additional

explanation in § 6.1)

Contrast

0%…100%

Adjustment of the LCD contrast

Time

Set real time (hour: minute)

Date

Set current date (day-month-year)

COM port

RS232 2400, RS232 4800,

RS232 9600,RS232 19200,

USB 115000

Set communication mode and rate

Language

Ita, Eng, Esp, Deu

Set system language

Initialization

For internal factory and service maintenance

only

Table 1: Configuration of system parameters


1. Use  and  arrows to select parameter (line) to be adjusted
2. Use  or  arrows to change the value of the selected parameter. If there are two or

more sub-parameters in one line (e.g. date and time) then use the SELECT key to skip
to the next sub-parameters and back

3. Press the ESC key to exit from configuration and back to the main menu