10 technical specifications, Insulation resistance measurement – HT instruments HT7052 User Manual
Page 36
![background image](/manuals/367514/36/background.png)
HT7052
EN - 34
10 TECHNICAL SPECIFICATIONS
Accuracy is given as [%rdg + (number of dgt) * resolution] at reference indicated in § 0
INSULATION RESISTANCE MEASUREMENT
Measurement range
Resolution
Accuracy
120k
999k 1k
(5.0rdg + 3dgt)
1.00M
9.99M 0.01M
10.0M
99.9M 0.1M
100M
999M 1M
1.00G
9.99G
0.01G
10.0G
99.9G 0.1G
100G
999G 1G
1.00T
10.00T 0.01T
(15.0 rdg + 3 dgt)
FS value of insulation resistance is defined as:
RFS = 1G
* Utest [V]
Nominal test voltage:
500
10kV DC
Current capability of test generator:
> 1mA
Short-circuit test current:
5mA
10%
Automatic discharge of tested object:
yes
Measurement range test voltage
Resolution
Accuracy
0
9999V
1V
(3.0 rdg + 3V)
10kV
0.1kV
3.0 rdg
Nominal test voltage:
500
10kV DC programmable in step of 25V
Accuracy of test voltage:
-0 / +10% + 20V
Output power:
10W max
Measurement range test current
Resolution
Accuracy
0.00
9.99nA
0.01nA
(5.0 rdg + 0.05nA)
10.0
99.9nA
0.1nA
100
999nA
1nA
1.00
9.99A 0.01A
10.0
9.99A 0.1A
100
999A 1A
1.00
5.50mA
0.01mA
Noise current rejection (resistive load)
Filter option
Maximum current @ 50Hz (mA rms)
Fil0 1.5
Fil1
2.5
Fil2
4.5
Fil3
5
Diagram Test voltage – Resistance
0
2
4
6
8
10
12
0,1
1
10
100
[ M ]
[ k
V
]
Utest=10kV
Utest=5kV