HT instruments HT7052 User Manual
Page 23
![background image](/manuals/367514/23/background.png)
HT7052
EN - 21
Fig. 24: Connection of instrument for insulation measurement
5. Press START/STOP key to activate the measurement in continuous mode
6. Wait for a stable result at display and press again START/STOP key to stop the
measurement or wait for the end of the set Timer. The result of test is shown at display
(see Fig. 21 or Fig. 23) with meaning of items descript in Table 4
7. Wait for the object under test to discharge
8. For saving the result see § 7
Parameter at display
Description
Fil0 (Fil1, Fil2, Fil3)
Filter type enabled, see the chapter 5.3. Configuration
(see § 6.3)
5000V
Set test voltage
U=5323V
Applied test voltage
I=266nA
Applied test current
19.9G
Result of insulation measurement
C=0.0nF Capacitance
of measured object
Tm:01min 04s
Timer information – test duration
Bargraph Analogue
representation of result
Rmax=20.G
Maximum value of result (only if timer is enabled)
Rmin=19.9G
Minimum value of result (only if timer is enabled)
Table 4: Meaning of parameters of insulation measurement
CAUTION
If the timer is disabled then OFF is displayed instead of the timer value
During a measurement, the timer information displays the time needed to
the complete the measurement (tr) while after the completion the test
duration (tm) is displayed
A high-voltage warning symbol appears on the display during the
measurement to warn the operator of a potentially dangerous test voltage
Value of capacitance is measured during the final discharge of the test
object