HT instruments HT7052 User Manual

Page 25

background image

HT7052

EN - 23

Fig. 27: Setting parameters

Fig. 28: Setting of graph R(t) parameters


6.6.2 Perform the measurement
1. Switch on the instrument by pressing the ON/OFF key
2. Select with arrow keys  or  the item “DIAGNOSTIC TEST” on main menu and

confirm with SELECT key. The screen of Fig. 29 is shown by the meter. In case of
activation of Graphic R(t) option press arrow keys  or  to select the graphical
screen of Fig. 31

CAUTION

It is not possible to switching mode of presentation when measurement
running

Fig. 29: Initial numerical screen

Fig. 30: Numerical screen of result

Fig. 31: Initial graphical screen

Fig. 32: Graphical screen of result


3. Connect the red part of the Test lead 1 or Test leads 2 (see § 4.2) to the +OUTPUT

input and the black part of the Test leads 2 (see § 4.2) to the –OUTPUT input. In case
of use of GUARD terminals (see § 6.2) connect also the green cables to the “GUARD”
input (see Fig. 33)

4. Connect the tip of Test lead 1 or Test leads 2 (positive) and the black cable of Test

leads 2 (negative) to the object on test (see Fig. 33)