HT instruments HT7052 User Manual

Page 37

background image

HT7052

EN - 35

DAR, PI, DD MEASUREMENT

Measurement range

Resolution

Accuracy

0.01

 9.99

0.01

(5.0 rdg + 2 dgt)

10.0

 100.0

0.1

5.0 dgt

Capacitance range for DD test:

5nF

 50F


INSULATION MEASUREMENT WITH STEP VOLTAGE

Measurement range test voltage

Resolution

Accuracy

2000

 9999V

1V

(3.0 rdg + 3V)

 10kV

0.1kV

3.0 rdg

Nominal test voltage:

2000

 10kV DC programmable in steps of 125V

Accuracy of test voltage:

-0 / +10% + 20V


WITHSTANDING VOLATGE DC

Measurement range test voltage

Resolution

Accuracy

500

 9999V

1V

(3.0 rdg + 3V)

 10kV

0.1kV

3.0 rdg

Measurement range leakage current

Resolution

Accuracy

0.000

 0.009mA

0.001mA

(3.0 rdg + 3 dgt)

0.01

 5.50mA

0.01mA

3.0 rdg

Nominal test voltage:

500

 10kV DC programmable in steps of 25V

Accuracy of test voltage:

-0 / +10% + 20V


AC or DC VOLTAGE

Measurement range

Resolution

Accuracy

0

 600V

1V

(3.0 rdg + 4V)

Input impedance:

3M

 10%

Voltage frequency

Resolution

Accuracy

0 e 45.0

 65.0Hz

0.1Hz

0.2Hz

Frequency within 0 and 45Hz:

displayed < 45Hz

Frequency > 65Hz:

displayed > 65Hz


CAPACITANCE

Measurement range

Resolution

Accuracy

0.0

 99.9nF

0.1nF

(5.0 rdg + 2 dgt)

100

 999nF

1nF

1.00

 50.0F 0.01F

Full-scale value of capacitance is defined as:

CFS = 10

F * Utest [kV]

10.1 SAFETY STANDARDS
Instrument safety:

IEC/EN61010-1, IEC/EN61557-2

Measuring accessory safety:

IEC/EN61010-031

Insulation: double

insulation

Protection:

IP44 (closed case)

Polluting level:

2

Overvoltage category:

CAT IV 600V (to earth), max 600V between inputs

Maximum altitude:

2000m (6561ft)