Caution – HT instruments HT7052 User Manual
Page 29
HT7052
EN - 27
5. Press START/STOP key to activate the insulation measurement
6. Wait for the end of the set Timers. The result of test is shown at display (see Fig. 39 or
Fig. 41) with meaning of items descript in Table 8
7. Wait for the object under test to discharge
8. For saving the result see § 7
Parameter at display
Description
Fil0 (Fil1, Fil2, Fil3)
Filter type enabled (see § 6.3)
5000V
Set test voltage – step 125 V
U=5302V
Applied test voltage
I=133nA
Applied test current
39.7G
Result of insulation measurement
C=0.0nF Capacitance
of measured object
Tm:05min 00s
Actual test duration
R1000V=40.0G
Last result of 1
st
step
R2000V=40.0G
Last result of 2
nd
step
R3000V=40.0G
Last result of 3
rd
step
R4000V=39.8G
Last result of 4
th
step
R5000V=39.7G
Last result of 5
th
step
U1=1076V 1
st
step voltage
U2=2141V 2
nd
step voltage
U3=3238V 3
rd
step voltage
U4=4278V 4
th
step voltage
U5=5302V 5
th
step voltage
Table 8: Meaning of parameters of insulation measurement with step voltage
CAUTION
Timer information is displayed from the start of the measurement until the
completion of each step measurement
A high-voltage warning symbol appears on the display during the
measurement to warn the operator of a potentially dangerous test voltage
The value of the capacitance is measured during the final discharge of the
test object