Fig. 33: Connection of instrument for diagnostic test
5. Press START/STOP key to activate the insulation measurement
6. Wait for the end of the set Timers. The result of test is shown at display (see Fig. 30 or
Fig. 32) with meaning of items descript in Table 6
7. Wait for the object under test to discharge
8. For saving the result see § 7
Parameter at display
Description
Fil0 (Fil1, Fil2, Fil3)
Filter type enabled on test (see § 6.3)
5000V
Set test voltage – step 25 V
U=5295V
Applied test voltage
I=55.6nA
Applied test current
95.1G
Result of insulation measurement
C=2.1nF Capacitance
of measured object
Bargraph Analogue
representation of Riso result
R15sec=95.0G
Resistance value measured after set time 1
R01min=95.1G
Resistance value measured after set time 2
R10min=95.1G
Resistance value measured after set time 3
DAR=1.00
DAR as ratio of R1min / R15s
PI=1.00
PI as ratio of R10min/R1min
DD= DD
result
Table 6: Meaning of parameters of diagnostic test
CAUTION
A high-voltage warning symbol appears on the display during the
measurement to warn the operator of a potentially dangerous test voltage
The value of the capacitance is measured during the final discharge of the
test object
If enabled, the instrument measures Dielectric Discharge (DD) when the
capacitance is in the range 5nF
50F