GW Instek GPI-700A Series HSB-001-01/-02 User manual User Manual
Page 7
SCANNER BOX HSB-001-1/2
USER MANUAL
8
5. Interconnect Cable
GHT-110
GHT-108
GHT-109
The GHT-110 control cable connects between the rear panel
SCAN connector of GPT/GPI-700A and the rear panel SCAN I/F
connector of the HSB-001-1/2.
The GHT-108 high voltage interconnect cable is then connected
from the high voltage rear output connector of the GPT/GPI-700A
to the rear high voltage input of the HSB-001-1/2.The rear panel
return of GPT/GPI-700A is connected to the rear panel input of
the HSB-001-1/2.
The GHT-109 high current interconnect cable is then connected
form the high current rear output connector of the GPI-745A to
the rear high current input of the HSB-001-1.
SETUP
Press MENU key to enter status MENU then use knob and arrow
keys to select a step.
Press EDIT/SAVE key to enter status EDIT. Now the cursor stays
at the “Test mode” field. Use the knob to select mode ACW, DCW,
IR and GB.
SCANNER BOX HSB-001-1/2
USER MANUAL
9
Functionality of FIELD key:
Test mode
ACW(DCW)
Output voltage
V=X.XXXkV
Measurement upper limit
Imax=XX.XXmA
Measurement lower limit
Imin=XX.XXmA
Ramping time
RAMP=XXX.Xs
Testing time
TEST=XXX.Xs
Channel
Hi=00
Lo=00