5. Interconnect Cable

GHT-110

GHT-108

GHT-109

The GHT-110 control cable connects between the rear panel

SCAN connector of GPT/GPI-700A and the rear panel SCAN I/F
connector of the HSB-001-1/2.

The GHT-108 high voltage interconnect cable is then connected

from the high voltage rear output connector of the GPT/GPI-700A
to the rear high voltage input of the HSB-001-1/2.The rear panel
return of GPT/GPI-700A is connected to the rear panel input of
the HSB-001-1/2.

The GHT-109 high current interconnect cable is then connected

form the high current rear output connector of the GPI-745A to
the rear high current input of the HSB-001-1.

SETUP
Press MENU key to enter status MENU then use knob and arrow
keys to select a step.
Press EDIT/SAVE key to enter status EDIT. Now the cursor stays
at the “Test mode” field. Use the knob to select mode ACW, DCW,
IR and GB.


SCANNER BOX HSB-001-1/2

USER MANUAL



9

Functionality of FIELD key:









Test mode

ACW(DCW)

Output voltage

V=X.XXXkV

Measurement upper limit

Imax=XX.XXmA

Measurement lower limit

Imin=XX.XXmA

Ramping time

RAMP=XXX.Xs

Testing time

TEST=XXX.Xs

Channel

Hi=00

Lo=00