DC MILLI-OHM METER

PROGRAMMER MANUAL

⎯ 52 ⎯

Queues

The output queue is included to power supplies.

Output Queue: The METER store query responses in the output queue by
succeeding the IEEE 488.2 protocol. If the METER receives a new
command or query message after a message terminator, the METER will
clear and reset this queue each time. The computer must read a query
response before it sends the next command (or query) or it loses response
to earlier queries.

Error/Event Queues
When an error or event occurs, the output queue stores the message. The
output queue stores and reports the messages on a FIFO (first in first out)
state. The SYSTem:ERRor? query reads the next item from the output
queue. If output queue overflows, the error message is –350, “Queue

overflow”

; the queue can’t store or report succeeding messages till it

is read or cleared.

Error Message

Table 8 lists the SCPI error messages for the METER.

Table 8 The error messages for the METER:

SCPI Error Code and Description

SESR Bit

0, “No error”
-100, “Command error”

5

-200, “Execution Error”

4

-221, “Settings conflict”

4

-222, “Data out of range”

4

-350, “Queue overflow”

3

-410, “Query INTERRUPTED”

2

-420, “Query UNTERMINATED”

2

-430, “Query DEADLOCKED”

2

DC MILLI-OHM METER

PROGRAMMER MANUAL