background image

DC MILLI-OHM METER

PROGRAMMER MANUAL

⎯ 30 ⎯

Examples:

:CONFigure:TCOMpensate:TEMPerature 20 sets require temperature to

20.0℃.

:CONFigure: TCOMpensate:TEMPerature? return “+2.0000E+1” to

indicate the 20.0℃.

The setting value and return values is in ℃ unit.

:CONFigure:SPEed

Function:

Set or query speed mode on or off.

Syntax:

:CONFigure:SPEed

can be o (off) or 1 (on).

:CONFigure:SPEed?

Return:

0/1

Examples:

:CONFigure:SPEed 1 sets speed mode, the conversation rate is 30
times/second.

:CONFigure:SPEed? return “1” to indicate the conversation rate is 30

times/second.

DC MILLI-OHM METER

PROGRAMMER MANUAL

⎯ 31 ⎯

:CONFigure:SCAN:STATe

Function:

Set or query scan mode on or off.

Syntax:

:CONFigure:SCAN:STATe

can be 0 (off) or 1 (on).

:CONFigure:SCAN:STATe?

Return:

0/1

Examples:

:CONFigure: SCAN:STATe 1 sets the scan mode on.

:CONFigure:SCAN:STATe? return “1” to indicate that the scan mode is

on.

:CONFigure:SCAN:CHANnel

Function:

Set or query the channel for scan.

Syntax:

:CONFigure:SCAN:CHANnel

is in the range from 0 to 1 .

:CONFigure:SCAN:CHANnel?

Return:

0/1