GW Instek GOM-802 Programming Manual User Manual
Page 11

DC MILLI-OHM METER
PROGRAMMER MANUAL
⎯ 20 ⎯
*WAI
Wait to continue. This command forces
sequential operation of commands. This
command is required by IEEE-488.1-1987.
The Meter, however, forces sequential
operation of commands by design.
:SYSTem:ERRor?
Read the next item from the error/event queue.
:SYSTem:VERSion?
Returns the SCPI version level.
*RCL
Recalls the setting data from the memory.
*SAV
Saves the setting data to memory.
DC MILLI-OHM METER
PROGRAMMER MANUAL
⎯ 21 ⎯
6. DETAILS OF COMMAND REFERENCE
Each command in this chapter will give a detailed description. The
examples of each command will be provided and what query form might
return.
*CLS
(no query form)
Function:
Clear all event status data register. This includes the Output Queue,
Operation Event Status Register, Questionable Event Status Register, and
Standard Event Status Register.
Syntax:
*CLS
Examples:
*CLS clears all event registers.
*ESE
Function:
Set or return the bits in the Event Status Enable Register (ESER). The
ESER enables the Standard Event Status Register (SESR) to be
summarized on bit 5 (ESB) of the Status Byte Register (SBR).
Syntax:
*ESE
*ESE?
Returns:
binary bits of the ESER.