beautypg.com

GW Instek GOM-802 Programming Manual User Manual

Page 17

background image

DC MILLI-OHM METER

PROGRAMMER MANUAL

⎯ 32 ⎯

Examples:

:CONFigure: :SCAN:CHANnel 8 sets the channel of scan.

:CONFigure:SCAN:CHANnel? Return 8 to indicate that there are 8

channels need to be sweeped.

:CONFigure:SCAN:DELay

Function:

Set or query the delay of scan.

Syntax:

:CONFigure:SCAN:DELay

is in the range from 30 to 30000. 16.2ms per unit.

:CONFigure:SCAN:DELay?

Return: is in the range from 30 to 30000.

Examples:

:CONFigure:SCAN:DELay 300 sets delay to 300 units..

:CONFigure:SCAN:DELay? Return “00300”, total 5 characters.

:CONFigure:BUZZer

Function:

Set or query the setting of the buzzer mode.

Syntax:

:CONFigure:BUZZer

can be 0 (on) 1(bp) or 2 (bf).

DC MILLI-OHM METER

PROGRAMMER MANUAL

⎯ 33 ⎯

:CONFigure:BUZZer?

Return:

0/1/2

Examples:

:CONFigure:BUZZer 0 sets the buzzer mode to be not active

:CONFigure:BUZZer? Return “1” to indicate that the buzzer mode is at

bp. (When the compare result comes out to be PASS, the buzzer will

sound the alert.)

:CONFigure:TRIGger

Function:

Set or query trig mode on or off.

Syntax:

:CONFigure:TRIGger

can be 0 (off) or 1 (on).

:CONFigure:TRIGger?

Return:

0/1

Examples:

:CONFigure:TRIGger 1 sets the trig mode on.

:CONFigure:TRIGger? Might return “1” to set the trig mode on