MEASUREMENT

17

1. If Trigger On was configured to

Data*, press Data from the bottom
menu.

*The applicable Trigger On settings
are shown below.

UART

Tx Data, Rx Data

I

2

C

Data, Address/Data

SPI

MOSI, MISO, MOSI&MISO

2. Press Number of Words/Number of

Bytes from the side menu and
choose data length.

UART

1~10 Bytes

I

2

C

1~5 Bytes

SPI

1~16 Words

3. If I

2

C is the bus type, press

Addressing Mode to toggle between
7 and 10 bit addressing modes.

4. Press Data (I

2

C, UART) or

MOSI/MISO (SPI) from the side
menu to edit the triggering data.

To edit the data, use the Variable
knob to highlight a binary or hex
digit and press Select. Use the
Variable knob to choose a value for
the digit and press Select to
confirm.

OR

Binary

0,1,X (don’t care)

Hex

0~F, X (don’t care)