Chapter 6

General Electrical Specifications

Repeatability +/- 1 Count (single channel only)

Stability with temperature

Zero: 1uV/

°

C

Span: .01%

Thermocouple Reference Junction Internal, automatic, 0.03
degrees C/degrees C, 5

°

C to 45

°

C.

Break Detection Upscale

50nA unit displays OL.

Stability with Time .8

°

/year

Noise Rejection

NMRR: 60dB @ 50/60 Hz
CMRR: 120dB @ 50/60 Hz (+/-0.1Hz with 250

unbalance)

Overload Protection

Power leads to ground: (1500VDC or VAC RMS)
Across inputs, for one minute:

T/C up to 250VDC or VAC, V+ to V-