Model CX-IR

CX-IR Bottom

Housing Assembly

Plug-in Replaceable

Sensor

Intelligent Transmitter Module (ITM)
Micro-processor controlled circuit

encapsulated in an Explosion proof
housing.

Magnetic

Programming Switches

Rain Shield

3

4

" NPT Locking

Nut

Ferrite Bead

Figure 3 Sensor Assembly Breakaway

NOTE

All metal components are constructed from electro polished 316 Stainless Steel to
maximize corrosion resistance in harsh environments.

CX-IR Intelligent Transmitter Module

The transmitter module is microprocessor-based and attached to the explosion proof junction box. Circuit
functions include an intrinsically safe barrier, on-board power supply, microprocessor, magnetic programming
switches, and a linear 4-20mA DC output. Magnetic program switches (located on either side of the ITM) are
activated by a hand-held magnetic programming tool, allowing non-intrusive operator interface with the
transmitter module. Electrical classifications are Class I, Div 1, Groups B, C, and D.









Figure 4 Functional Block Diagram

Field Replaceable Sensor
Detconꞌs infrared gas sensors are field proven, plug-in sensors with over-sized gold-plated connections that
eliminate corrosion problems. The sensor can be accessed and replaced in the field easily by releasing the
locking screw and unthreading the splashguard adapter assembly.

Figure 5 Plug-in Sensor

I/O

Circuit

Protection

mA Output

Power In

Power

supplies

Micro-

Processor

Intrinsically

Safe Barrier


Plug-In

Sensor

Modbus Output

CX-IR Instruction Manual