From the Set Modbus ID scroll, hold the programming magnet over PGM1 or PGM2 until the "

►" prompt

appears and continue to hold the magnet in place for an additional 3 to 4 seconds (until the display starts to
scroll Set ID). The display will then switch to XX (where XX is the current ID address).

Swipe the magnet over PGM2 to increase or PGM1 to decrease the hexadecimal number until the desired ID
is displayed. Hold the magnet over PGM2 for 3 to 4 seconds to accept the new value. The display will scroll
ID Saved, and revert to Set Modbus ID text scroll.

Move to another menu item by executing a momentary hold or, return to normal operation by automatic
timeout of about 15 seconds (the display will scroll Set Serial ID 5 times and return to normal operation).

3.5 Fault Diagnostic/Failsafe Feature

If the ITM should incur a fault, the Global Fault bit will be set on the Modbus™ output. This can occur if the
ITM detects a problem with the sensor, detects that there is no sensor connected, if the ITM has an internal
fault, or other fault condition. The Global Fault bit will be set on the Modbus™ output until the problem is
resolved. The display will show the Fault when a magnetic programming tool is swiped across either PGM1 or
PGM2. The error codes are defined in Section 6 Troubleshooting.