Model DM-700

DM-700 Instruction Manual

Rev. 4.0

Page 37 of 48

9. Appendix

9.1 Specifications

System Specifications

Sensor Type:

Continuous diffusion/adsorption type
3-Electrode Electrochemical Sensor (2-Electrode for O

2

)

Plug-in Replaceable Intelligent Type

Sensor Life:

2 years typical

Measuring Ranges:

0-1 ppm up to 0-10,000 ppm (Toxic Gasses)
0-100 ppm up to 0-25% volume (O

2

)

Accuracy/ Repeatability:

±10% of applied gas or +/- 2ppm, greater of (ISA 92.0.01 tested for H

2

S)

±2% of full-range (other Toxic Gasses)
±1% of full-range (O

2

)

Response Time:

T90 < 30 seconds typical (See Sensor Table)

Performance Testing:

Complies with ISA 92.0.01 Part 1-1998; Performance Requirements for H

2

S

Electrical Classification:

CSA and US (NRTL)
Class I, Division 1, Groups A, B, C, D

ATEX
Class I, Zone 1, Group IIC
II 2 G Ex d [ib] ib IIC T4 Gb

Approvals:

C

CSA

US

, ATEX, CE Marking

Warranty:

Electronics – 2 years
Sensor – (See Table 4)

Environmental Specifications

Operating Temperature:

-40°C to +50°C typical (See Table 4)
(ISA 92.0.01 tested from -40°C to +50°C for H

2

S)

Storage Temperature:

-35°C to +55°C typical

Operating Humidity:

10-95% RH Continuous Duty (See Table 4)
(ISA 92.0.01 tested to 5-95% RH for H

2

S)

0-100% RH Short-Term Duration Only

Operating Pressure:

Ambient ± 10%

Air Velocity:

0-5 meters/second