beautypg.com

Detcon DM-700 User Manual

Page 37

background image

Model DM-700

DM-700 Instruction Manual

Rev. 4.0

Page 33 of 48

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses

 If using Splashguard with Integral Cal Port, must use Calibration Wind Guard or air movement can

compromise span gas delivery.

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life

 Replace the plug-in toxic sensor.

Stability Fault

Probable Causes: Failed Sensor, empty or close to empty Cal Gas Cylinder, problems w/ cal gas and delivery.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses

 If using Splashguard with Integral Cal Port, must use Calibration Wind Guard or air movement can

compromise span gas delivery.

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life

 Replace the plug-in toxic sensor.

Clearing Fault

Probable Causes: Failed Sensor, Cal Gas not removed at appropriate time, problems w/ cal gas and delivery,
Background of Target Gas.

 Must recover to < 5% of range in < 5 min after AutoSpan is complete

 Use bottled air (zero air or N

2

) if there is a known continuous background level.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life

 Replace the plug-in toxic sensor.

Poor Calibration Repeatability

Probable Causes: Failed Sensor, use of wrong Cal Gas or problems w/ cal gas and delivery, Interference
Gasses.

 Check for adequate Sensor Life.

 Check validity of span gas using pull tube or other means (check MFG date on cal gas cylinder).

 Use proper cal gas regulators and tubing for highly corrosive gasses

 Check for obstructions affecting cal gas hitting sensor face (including being wet, blocked, or

corroded).

 Verify adequate Sensor Life

 Replace the plug-in toxic sensor.

Unstable Output/ Sudden spiking

Possible Causes: Unstable power supply, inadequate grounding, or inadequate RFI protection.