6. Troubleshooting Guide

Refer to the list of Failsafe Diagnostic features listed in Section 3.6.2 for additional reference in
troubleshooting activities. Listed below are some typical trouble conditions and their probable cause and
resolution path.

ITM

Bottom View

Display

Window

Plug in Sensor

Bottom view

Locking

Setscrew

Locking

Setscrew

Place programming

connector towards

back away from

display window

PLUG-IN O
ALIGN AR
WITH LED

ntents on skin

MODEL

DM-700

detcon inc.

H

2

S Sensor

detcon inc.

Figure 16 Sensor Cell and ITM Mating

Under-Range Fault

Probable Cause: Sensor Baseline drifted lower, Interference gasses.

 Repeat AutoZero. Use Zero Air or N

2

source.

 Allow more time for zero stabilization if this is a biased sensor type.

 Execute successful AutoSpan and verify adequate Sensor Life.

 Check Raw counts in View Sensor Status. Should be close to 33,000 counts when normal.

 Replace plug-in sensor if fault continues.

Missing Sensor Fault

Probable Cause: Sensor is Missing, Failed Plug-in Sensor Electronics, or ITM I.S. Barrier Failure.

 Make sure plug-in sensor is plugged in properly with correct orientation.

 Swap plug-in sensor into another ITM to determine if plug-in sensor problem or ITM problem

 Replace the plug-in sensor if proven faulty

 Replace the ITM if proven faulty

AutoSpan Calibration Faults – (Range, Stability and Clearing)

To clear any AutoSpan Calibration fault, the AutoSpan process must be completed successfully (Section3.4).

Range Fault

Probable Causes: Failed Sensor, Cal Gas not applied or not applied at appropriate time, problems w/ cal gas
and delivery.